When a validated linear standard curve for one analyte is already known, enter its slope and intercept plus a measured sample signal. The calculator solves the line for concentration and reverses a separately entered dilution factor.

Enter your values

Result

Results are based only on the inputs and assumptions shown here. Nothing you enter is sent or saved.

Formula: measured aliquot mg/mL = (sample signal − intercept) ÷ slope; original sample mg/mL = measured aliquot mg/mL × dilution factor

Worked example

If signal is 0.52, slope is 0.10 signal per mg/mL and intercept is 0.02, the measured aliquot is 5 mg/mL. With a documented fivefold dilution, the original sample is 25 mg/mL.

A curve must already exist for this analyte

This page does not fit standards or determine whether a calibration is linear. Use a curve with a validated response range and the same signal and concentration units. NIST calibration material describes solving a straight-line signal relationship for concentration.

Dilution factor is explicit

Enter 1 for an undiluted measurement. A fivefold dilution means the measured aliquot concentration is one fifth of the original under a conserved-analyte assumption. Verify any actual preparation and recovery losses outside this tool.

Interpretation limits

A signal below the entered intercept is rejected rather than reported as a negative concentration. An in-range result still does not establish chemical identity, assay quality, precision, clinical suitability or a dose. Arithmetic stays in the browser.